For as long as patients have been intubated, the potential for unplanned removal of the endotracheal tube has been a source of concern. This concern has led nurses and other medical professionals to search for ways to decrease the occurrence of unplanned extubation. Unplanned extubation is defined as any accidental or purposeful removal of the endotracheal tube by a patient.13
Because of a high rate of unplanned extubations2.14% in the medical intensive care unit (MICU)/coronary care unit (CCU) and 2.32% in the surgical ICU (SICU)at a tertiary care hospital in north central Wisconsin, a quality control committee was formed to address methods for decreasing the unplanned extubation rate and increasing awareness of the rate of unplanned extubations. The committee consisted of staff from respiratory therapy, clinical nursing, and nursing administration and a physician champion. As information and data were evaluated, the committee determined that although sedation and restraint were intermittently contributing factors to unplanned extubation, the techniques used for securing endotracheal tubes were the one factor that was consistently a problem. Staff members were using several different techniques to secure patients endotracheal tubes. Some patients had tube holders from other facilities that were unfamiliar to the staff; other patients had different methods of taping and products used. Some patients had their tube secured around their head; others had tape adhered to their bare skin.
Because of the identified inconsistencies, the committee developed a plan to educate the nursing and respiratory care staff. The education consisted of training in new securing techniques, heightening awareness of patients activity levels, and improving knowledge about patients readiness for extubation. In this article, we address issues associated with unplanned extubations and how to decrease the occurrence of unplanned extubation by using a consistent securing technique.
| Literature Review |
|---|
|
|
|---|
Quality improvement and education both played an important role in significantly decreasing rates of unplanned extubation.11,12 Although some authors discussed new methods of securing endotracheal tubes, none provided information about the processes used to instruct staff in using the new methods properly. Whether securing techniques actually decrease the frequency of unplanned extubation in critical care populations also was not discussed.
| Methods |
|---|
|
|
|---|
An intensivist, the respiratory care staff, and the nursing staff all perceived a high frequency of unplanned extubations in the adult critical care units during a 6-month period. The policy for securing endotracheal tubes was due for an update. Additionally, the products and techniques used to secure endotracheal tubes had changed, and those changes had not been incorporated into the policy and procedure. Therefore, staff members did not know how to use and secure endotracheal tubes properly.
The committee thought that extubations could have been avoided if different techniques and possibly different products had been used to secure endotracheal tubes. To test this theory, the quality control committee developed a tool to track factors related to unplanned extubations (Figure 1
). For each unplanned extubation, nurses in the critical care units were asked to enter information on the form. They were also required to complete an incident report, which was submitted to the quality control group and the risk management department at the hospital.
|
Upon reviewing the data, the committee deemed the frequency of unplanned extubations to be too high, 2.32% in the SICU and 2.14% in the CCU/MICU. A review of the literature did not reveal any nationally accepted standard. Therefore, a goal was set to reduce the frequency of unplanned extubations to less than 0.8%; the committee thought that doing so would indicate a higher quality of care. To meet this goal, the quality control group investigated different tube holders in an effort to improve the security of the endotracheal tube. Evaluation of the tube holders required staff to receive instruction to ensure proper use of each holder.
When this investigation was done, several manufacturers were producing endotracheal tube holders. The committee examined each of the holders available, looking at design, securability, ease of application, and durability once in place. They also wanted a device that allowed easy movement of the tube from side to side in order to maintain the integrity of the oral mucosa. Another feature important to the committee was that the holder maintain skin integrity beneath the area of application. This evaluation helped the committee narrow the field of prospective devices to 2, 1 (Hollister oral endotracheal attachment device) manufactured by Hollister, Inc (Libertyville, Ill) and 1 (Comfit) manufactured by Ackrad Laboratories, a CooperSurgical Company (Cranford, NJ). These 2 types of holders were tried by the staff of each unit.
After determining which devices were to be evaluated, the committee created a teaching plan to prepare the nursing and respiratory staff to use the new devices. A new policy that included instruction in using each device was created and put into use. The existing policy and procedure for taping were updated to provide an alternative choice for securing endotracheal tubes. Hands-on instruction about each holder and the taping technique was provided to all staff members along with written information and posters demonstrating proper use of the tube holders. Instruction was offered for 2 weeks, with times dispersed over all shifts to facilitate staff compliance.
During the in-service training sessions, the nursing and respiratory care staff were taught the manufacturers recommended application technique for the 2 tube holders. The staff were also taught a method of taping, using cloth tape circling behind the head and secured directly to the tube. A piece of tape was torn to a length that extends around the back of the patients head, with adhesive ends long enough to wrap around the endotracheal tube and make the tube secure. An adhesive skin preparation was applied to the cheek area and the endotracheal tube to give a sticky base to which the tape could adhere. The upper loose ends were then wrapped around the endotracheal tube with the lower loose ends being brought up, over, and across the patients upper lip to the opposite cheek. The current policy and procedure for using tape as a securing device was updated, and a new policy was created to describe the use of the new tube holder in order to maintain consistency among staff members when choosing and using securing devices.
Once the training was complete, each device was evaluated for separate 3-month periods to allow for an application learning curve to level out. With each evaluation, the education provided for that securing device was reinforced.
| Results |
|---|
|
|
|---|
During this project, the unplanned extubation rates for 2 6-month periods were compared to determine whether the frequency of unplanned extubations decreased after extensive changes in the process of securing endotracheal tubes were coupled with intensive education about the changes. When the first 6 months preceding staff education was compared with the 6 months after staff education, an improvement in the rate of unplanned extubation was apparent among all patients studied (Figure 2
).
|
|
| Discussion |
|---|
|
|
|---|
| Limitations of the Study |
|---|
|
|
|---|
Another limitation of this study is that the committee did not use a formal database to collate information but rather tallied all results by hand. Because of this practice, only patients who had an unplanned extubation were followed up; therefore, patients having planned extubations could not be compared with patients who had an unplanned extubation. If a formal database had been used, all intubated patients might have been included in order to compare some of the interventions that were used successfully for patients who stayed intubated versus patients who had unplanned extubations. Sedation and/or use of restraints, oral versus nasal intubation, and the need for reintubation after unplanned extubation also could have been compared in patients who remained intubated versus patients with unplanned extubations. Use of a formal database would have kept information together, with better accessibility for future comparisons and research.
| Conclusion |
|---|
|
|
|---|
In this project, a perceived problem was assessed, and a tool for data collection was created. The tool was used to collect data on unplanned extubation and was then used to clarify perceptions and assumptions, steps that led to a change in practice. This change was of great benefit to the patients and the staff, because it provided consistency for a method of securing endotracheal tubes, which led to better quality of care for the patients. The tool has been used consistently for some time. We plan to try to build a database in which information can be stored and collated to allow comparison of methods and types of sedation and to develop protocols for fast-track weaning parameters and alternatives for restraint; our hope is to further decrease the number of unplanned extubations.