Critical Care Nurse. 2004;24: 67-70
Copyright © 2004 by the American Association of Critical-Care Nurses.
This Article
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Keckeisen, M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Keckeisen, M.


Protocols for Practice

Monitoring Pulmonary Artery Pressure

Maureen Keckeisen, RN, MSN, CCRN


Maureen Keckeisen is a clinical nurse specialist at UCLA Medical Center and assistant clinical professor at the UCLA School of Nursing in Los Angeles, Calif.

To purchase reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. Phone, (800) 809-2273 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints{at}aacn.org.

This article is based on the protocol Pulmonary Artery Pressure Monitoring by Maureen Keckeisen. It was taken from the Hemodynamic Monitoring series of AACN’s Protocols for Practice. Protocols can be obtained from AACN, 101Columbia, Aliso Viejo, CA 92656-1491, (800) 899-AACN, (949) 362-2000. $11, AACN members; $14, nonmembers.


Q: What is the proper reference point to use when leveling (referencing) and zeroing hemodynamic monitoring systems?

The phlebostatic axis provides an external reference point that approximates the anatomic level of the left and right atria and the pulmonary artery. Leveling (referencing) and zeroing the hemodynamic monitoring system to the tip of the catheter that lies within the pulmonary artery ensures that hemodynamic values obtained with the catheter are accurate.1 The air-fluid interface (zeroing stopcock), not the transducer, should be used when the hemostatic monitoring system is being leveled (referenced) to the phlebostatic axis. Recent research with computed tomography has confirmed that the phlebostatic axis approximates the level of the left atrium at the point midway between the anterior and posterior surfaces of the chest at the fourth intercostal space2,3 when the patient is supine (Figure 1Go).



View larger version (27K):
[in this window]
[in a new window]
 
Figure 1 Referencing and zeroing the hemodynamic monitoring system in a supine patient. The phlebostatic axis is determined by drawing an imaginary vertical line from the fourth intercostal space at the sternal border to the right side of the chest (A). A secondary imaginary line is drawn horizontally at the level of the midpoint between the anterior and posterior surfaces of the chest (B). The phlebostatic axis is located at the intersection of points A and B.

 
Q: What is the best position in which to place the patient to ensure accuracy when monitoring pulmonary artery pressure?

A variety of backrest positions may be used that do not compromise the accuracy of measurements of pulmonary artery pressure. Back-rest elevations of 0°, 30°, 45°, and 60° with the patient supine (back flat against bed surface) do not affect the accuracy of measurements of pulmonary artery pressure if the hemodynamic monitoring system has been properly referenced and zeroed.4–10 As the patient moves from flat to higher levels of backrest elevation, the reference level must remain horizontal to the phlebostatic axis in order for measurements of pulmonary artery pressure to remain accurate (Figure 2Go).



View larger version (39K):
[in this window]
[in a new window]
 
Figure 2 The level of the phlebostatic axis as the patient moves from flat to higher levels of back-rest. The level of the axis for referencing and zeroing the air-fluid interface rotates on the axis and remains horizontal as the patient moves from flat to increasingly higher backrest positions. For accurate hemodynamic pressure readings at different backrest elevations, the air-fluid interface must be at the level of the phlebostatic axis.

 
Q: Are pulmonary artery pressures measured in patients who are positioned on the side (lateral or side-lying position) accurate?

Monitoring pulmonary artery pressures in patients who are in lateral or side-lying positions (eg, 20°, 30°, 45° to the side) is generally not recommended. Most research results are conflicting and inconclusive because of the lack of a standardized point of reference for leveling and zeroing in patients in side-lying positions.11–18 The only lateral position for which measurements of pulmonary artery pressure seem to be accurate is with patients lying on the side in a 90° lateral position with the backrest flat.3 The landmarks for leveling and zeroing with patients in the right and left lateral positions areas follows: (a) right lateral position (at 90° with backrest flat)—the intersection of the fourth intercostal space and the midsternum; (b) left lateral position (at 90° with backrest flat)—the intersection of the fourth intercostal space and the left parasternal border. Because most critically ill patients who require monitoring of pulmonary artery pressure are rarely positioned in a 90° side-lying position, until external landmarks for the correct reference location in 20° to 45° side-lying positions are determined by further research, pulmonary artery pressure should not be monitored while patients are lying on the side (Figure 3Go).



View larger version (23K):
[in this window]
[in a new window]
 
Figure 3 Referencing and zeroing the hemodynamic monitoring system in a patient in a lateral position. A, For the right lateral position, the reference point is at the intersection of the fourth intercostal space and the midsternum. B, For the left lateral position, the reference point is the intersection of the fourth intercostal space and the left parasternal border.

 
Q: Is it normal for measurements of pulmonary artery pressure to fluctuate?

Fluctuations in pulmonary artery pressure are normal in critically ill adults but should be assessed for individual variations and clinical significance. Limited clinical studies indicate that normal fluctuations in pulmonary artery pressure of 4 mm Hg occur for pulmonary artery diastolic pressure and pulmonary artery wedge pressure (PAWP), and normal fluctuations of 5 mm Hg occur for pulmonary artery systolic pressure.19

Q: Are measurements of pulmonary artery pressure accurate in patients being treated with positive end-expiratory pressure (PEEP) during mechanical ventilation?

For patients being treated with PEEP of less than 10 cm H2O, correlation between the PAWP and left atrial pressure is good, in the absence of hypovolemia. Accuracy also depends upon having the tip of the pulmonary artery catheter below the level of the left atrium or lung zone 3 (dependent blood flow). If the tip of the catheter is in lung zone 1 (upper lung, no blood flow) or lung zone 2 (intermittent blood flow), the PAWP may reflect alveolar airway pressure rather than vascular pressure, resulting in a falsely elevated PAWP20–25 (Figure 4Go). For patients being treated with PEEP greater than 10 cm H2O, the accuracy of measurements of pulmonary artery pressure may be compromised, and data should be interpreted cautiously. PEEP greater than 10 cm H2O increases alveolar and intrathoracic pressure, compresses the pulmonary vasculature, and affects the accuracy of PAWP measurement. Graphic recording of the hemodynamic waveform is essential for accurate measurement of PAWP, particularly if respiratory variation is significant and the patient is being treated with high levels of PEEP.23,24,26–28 The accuracy of measurements of pulmonary artery pressure should be questioned if one of the following occurs:



View larger version (55K):
[in this window]
[in a new window]
 
Figure 4 Physiologic lung zones. The accuracy of PAWP as a reflection of pressure in the LA depends on having the catheter tip positioned below the level of the LA or in lung zone 3. Zone 1 has no blood flow, and alveolar pressure exceeds arterial pressure (PA Pa Pv). In zone 2, blood flow is intermittent, and arterial pressure exceeds alveolar and venous pressure (PaPA Pv). In zone 3, both arterial and venous pressures exceed alveolar pressure (Pa Pv PA), and PAWP reflects vascular pressures rather than alveolar pressures.

 

The following correction calculation can be used to estimate PAWP and correct for high levels of PEEP27:

  1. Convert the applied PEEP from centimeters of water to millimeters of mercury (1.36 cm H2O =1 mm Hg).
  2. Subtract half the applied PEEP in millimeters of mercury from the measured PAWP.

For example, in a patient with a measured PAWP of 24 mm Hg who is receiving 16 cm H2O applied PEEP, the correction calculation would be as follows:

Q: Which is the most accurate and reliable method of measuring pulmonary artery pressure, graphic recording or digital data taken directly from the monitor?

Studies have shown that measurements of pulmonary artery pressure are more accurate and reliable when taken from a graphic strip recorder than when based on digital data taken directly from the monitor. Use of graphic strip recordings enhances detection of phases of respiration and makes determination of end-expiration much more accurate. These factors are particularly important if respiratory variation is significant or high PEEP is used in a patient receiving mechanical ventilation. Significant discrepancies in measurements of pulmonary artery pressure can occur, resulting in falsely high measurements inpatients receiving mechanical ventilation and falsely low pressures in spontaneously breathing patients when digital data are taken directly from the monitor.29–31

Note

This article was first published in Critical Care Nurse December 1999.

References

  1. Winsor T, Burch GE. Phlebostatic axis and phlebostatic level: reference levels for venous pressure measurements in man. Proc Soc Exp Biol Med. 1945;58:165–169.
  2. Paolella LP, Dorfman GS, Cronan JJ, Hasan FM. Topographic location of the left atrium by computed tomography: reducing pulmonary artery calibration error. Crit Care Med. 1988;16:1154–1156.[Medline]
  3. Kee LL, Simonson JS, Stotts NA, Skov P, Schiller NB. Echocardiographic determination of valid zero reference levels in supine and lateral positions. Am J Crit Care. 1993;2:72–80.
  4. Cenzig M, Crapo RO, Gardner RM. The effect of ventilation on the accuracy of pulmonary artery and wedge pressures. Crit Care Med. 1983;11:502–507.[Medline]
  5. Chulay M, Miller T. The effect of backrest elevation on pulmonary artery and pulmonary capillary wedge pressures in patients after cardiac surgery. Heart Lung. 1984;13:138–140.[Medline]
  6. Clochesy JM, Hinshaw AS, Otto CW.Effect of change of position on PA and PCW pressures in mechanically ventilated patients. NITA. 1984;7:223–225.[Medline]
  7. Laulive JL. Pulmonary artery pressures and position changes in the critically ill adult. Dimes Crit Care Nurse. 1982;1:28–34.
  8. Wilson AE, Birmingham-Mitchell K, Wells N, Zachary K. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264–270.
  9. Woods SLY, Mansfield LT. Effect of body position upon pulmonary artery and pulmonary capillary wedge pressures in noncritically ill patients. Heart Lung. 1976;5:83–90.[Medline]
  10. Woods SLY, Grose BL, Laurent-Bopp D. Effect of backrest position on pulmonary artery pressures in critically ill patients. Cardiovasc Nurse. 1982;18:19–24.
  11. Cason CL, Holland LC, Lambert CW, Huntsman KT. Effects of backrest elevation and position on pulmonary artery pressures. Cardiovasc Nurse. 1990;26:1–6.
  12. Groom L, Frisch SR, Elliot M. Reproducibility and accuracy of pulmonary artery pressure measurement in supine and lateral positions. Heart Lung. 1990;19:147–151.[Medline]
  13. Keating D, Bolyard K, Eichler E, Reed JIII. Effect of side lying positions on pulmonary artery pressures. Heart Lung. 1986;15:605–610.[Medline]
  14. Kennedy GT, Bryant A, Crawford MH.The effects of lateral body positioning on measurements of pulmonary artery and pulmonary wedge pressures. Heart Lung. 1984;13:155–158.[Medline]
  15. Murphy CE. The Effect of Lateral Positioning on Pulmonary Artery and pulmonary Capillary Wedge Pressure in Critically Ill Patients [master’s thesis]. Eugene, Ore: University of Oregon; 1977.
  16. Osika CA. Measurement of pulmonary artery pressures: supine versus side lying head-elevated positions [abstract]. Heart Lung. 1989;18:298–299.
  17. Whitman G. Comparison of pulmonary artery catheter measurements in 20 degree supine and 20 degree right and left lateral recumbent positions [abstract]. Heart Lung. 1982;11:256–257.
  18. Wild L. Effect of lateral recumbent posittons on measurement of pulmonary artery and pulmonary capillary wedge pressures in critically ill adults [abstract]. Heart Lung. 1984;13:305.
  19. Nemens EJ, Woods SLY. Normal fluctuations in pulmonary artery and pulmonary capillary wedge pressures in acutely ill patients. Heart Lung. 1982;11:393–398.[Medline]
  20. Bridges EJ, Woods SLY. Pulmonary artery pressure measurement: state of the art. Heart Lung. 1993;22:99–111.[Medline]
  21. Jardin R, Farcot JC, Boisante L, et al. Influence of positive end-expiratory pressures on left ventricular performance. N Engl J Med. 1981;304:387–392.[Abstract]
  22. Lozman J, Powers SR, Older T, et al. Correlation of pulmonary wedge and left atrial pressures. Arch Surg. 1974;109:270–277.[Abstract/Free Full Text]
  23. Rajacich N, Burchard KW, Hasan RM, Singh AK. Central venous pressure and pulmonary capillary wedge pressure as estimates of left atrial pressure: effects of positive end-expiratory pressures and catheter tip malposition. Crit Care Med. 1989;17:7–11.[Medline]
  24. Shasby DM, Dauber IM, Pfister S. Swan-Ganz catheter location and left atrial pressure determine the accuracy of wedge pressure when positive end-expiratory pressure is used. Chest. 1981;80:666–670.[Abstract/Free Full Text]
  25. West JB, Dollery CT, Naimark A. Distribution of blood flow in isolated lung: relation to vascular and alveolar pressures. J Appl Physiol. 1964:19:713–724.[Abstract/Free Full Text]
  26. Gardner P. Pulmonary artery pressure monitoring. AACN Clin Issues. 1993;4:98–119.
  27. Marini JJ, O’Quin R, Culver BH, Butler J. Estimation of transmural cardiac pressure during ventilation with PEEP. J Appl Physiol. 1982;53:384–391.[Abstract/Free Full Text]
  28. Pinsky M, Vince JL, Desmet JM. Estimating left ventricular filling pressures during positive end-expiratory pressure in humans. Am Rev Respir Dis. 1991;143:25–31.[Medline]
  29. Maran AG. Variables in pulmonary capillary wedge pressure: variation within intrathoracic pressure, graphic and digital recorders. Crit Care Med. 1980;8:102–105.[Medline]
  30. Johnson MK, Schumann L. Comparison of three methods of measurement of pulmonary artery catheter readings in critically ill patients. Am J Crit Care. 1995;4:300–307.
  31. Wild LR, Woods SLY. Comparison of three methods for interpreting pulmonary artery wedge pressure waveforms with respiratory variation. Heart Lung. 1985;14:308–309.




This Article
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Keckeisen, M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Keckeisen, M.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS