Critical Care Nurse. 2005;25: 44-51
Copyright © 2005 by the American Association of Critical-Care Nurses.
Clinical Article
Indwelling Foley Catheters
Is the Current Design a Source of Erroneous Measurement of Urine Output?
Wendy M. Fallis, RN, PhD
Wendy M. Fallis is the director of research and evaluation at Victoria General Hospital, Winnipeg, Manitoba. Previously she was the clinical instructor in the intensive and postanesthetic care units. Her area of research is thermoregulation and temperature measurement, with a focus on adult critical care patients. She is a member of the research committee of the Canadian Association of Critical Care Nurses and the recipient of the Sigma Theta Tau 2003 Research Dissertation Award. The concerns expressed in this article arose during her doctoral research at the University of Washington in Seattle.
To purchase reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. Phone, (800) 809-2273 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints{at}aacn.org.
A frequent task undertaken by critical care nurses is the measurement of urine output. In the intensive care unit (ICU), hourly measurements of urine output are a routine procedure for critically ill patients. The amount of urine formed each hour is often used as an indicator of fluid status and renal function, and a decrease in the volume to less than a critical level is a matter of concern.
The task of measuring urine output is simplified when an indwelling urinary catheter is in place. For critically ill patients, insertion of a urinary catheter is warranted,1 and continued catheterization is justified when monitoring of urinary output is required.2 Accurate measurement of the rate of urine flow, however, may be hampered by the current design of indwelling catheters. In this article, I describe potential problems associated with the current design of indwelling bladder catheters and encourage the redesign by manufacturers of these catheters in North America to assist in the complete emptying of the bladder.
 |
Urine Production
|
|---|
Urine production is the result of continuous filtration of plasma through the kidneys, which receive approximately 20% to 25% of the cardiac output. Normally, less than 1% or approximately 1500 mL of the 180 L of plasma filtered daily leaves the body as urine. However, this quantity can vary from 600 to 2500 mL/d. Urine production generally occurs at a rate of 1 to 2 mL/min but can reach a maximum of 20 mL/min and depends on several factors. Urine volume has a direct relationship with fluid intake and the rate of renal circulation and an inverse relationship with the amount of water lost through other routes. In addition, the volume of urine depends on hydrostatic (blood) pressure; the number of active glomeruli; the permeability of the glomerular membrane; tubule reabsorption, which depends on antidiuretic hormone; and the osmotic pressure within the renal tubules.
Urine output is reduced when patients are hyperventilating or feverish or lose fluid via vomiting, a nasogastric tube, burns, or wounds. Output also can be decreased as a result of reduced renal blood flow and glomerular filtration, as occurs in patients undergoing surgery with cardiopulmonary bypass. Low cardiac output for these patients during the perioperative period can lead to renal dysfunction 12 to 18 hours after surgery.3 Depressed renal function is also a concern postoperatively in elderly patients4 and after traumatic injury.5 An adequate urine flow is greater than 500 mL/24 h per square meter or greater than 20 mL/h per square meter.3 In critical care, accurate hourly measurements of urine output are thus essential for evaluation of fluid status and renal perfusion.6
 |
Urine Transport and the Urinary Bladder
|
|---|
The bladder is the bodys repository for storage of urine. This highly compliant receptacle generally stores 500 mL of urine or more and is composed of an adventitial layer of soft connective tissue, an intermediate nonstriated smooth muscle coat (detrusor muscle), and an interior lining of mucous membrane. Upon filling, the bladder distends, and its walls are stretched, resulting in a thinning of the layers of transitional epithelium as the layers slide past each other.7
Situated behind the symphysis pubis (Figure 1
),8 the bladder lies entirely in the lower part of the pelvic cavity when empty and expands into the abdominal cavity upon distension. Its position also varies according to sex; it lies in front of the vagina and the uterus in females and anterior to the rectum in males. Various geometric shapes of human urinary bladders have been noted.9

View larger version (34K):
[in this window]
[in a new window]
|
Figure 1 Position of urinary bladder in men and women and tenting of urethral catheter up the dome of the bladder.
|
|
Urine formed in each kidney is propelled into the bladder via the ureters. These structures, each approximately 4 to 5 mm in diameter and 27 to 30 cm long,10 are composed of an external adventitial layer, a thick intermediate nonstriated muscle layer, and an inner mucosal layer that is extensively folded. The ureters lie beneath the parietal peritoneum and follow the posterior part of the abdominal wall to the pelvic brim, where they follow the lateral wall of the pelvis and the pelvic floor to enter the posterior part of the bladder wall. Peristaltic activity, influenced by urine volume, moves urine along the ureter into the bladder in boluses. These peristaltic waves occur at a limited rate of approximately 1 to 6 per minute.11 Because of the limited peristaltic rate, increased flow rates are accommodated by the formation of larger urine boluses within the ureters.11 A sphincterlike muscular fold guards the opening between the ureter and the bladder and opens with the arrival of a peristaltic wave.
Urine exits the body from the inferior side of the bladder via the urethra, which is considerably shorter in females (34 cm) than in males (1820 cm). In most persons, complete or almost complete emptying of the bladder occurs with each void and generally less than 10 mL of urine remains.12
 |
Indwelling Urinary Bladder Catheters
|
|---|
Approximately 10% of all hospitalized and long-term care patients in the United States require an indwelling urethral catheter.13 Bladder catheterization via the urethra is the most frequent retrograde procedure performed on the urinary tract14 and is a common occurrence for many patients undergoing surgery with general anesthetics, including cardiac surgery patients. In critically ill patients, urethral catheterization is used to assess urinary output.14 A 14F, 16F, or 18F catheter with a 5-mL balloon is recommended in adults,15 although a 16F catheter is commonly used.13 Dividing the Charrière French size by 3 converts it to millimeters and gives the external diameter of the catheter. Upon inflation, the balloon near the tip of the catheter maintains a retention catheter within the bladder. Gentle tugging on the tube immediately after insertion and inflation of the balloon ensures that the catheter is in the urethral orifice.16 A larger volume of fluid than that stated on the balloon port is required to account for filling of the inflation lumen and valve arm.17 For example, 10 mL of sterile water is the recommended amount for inflating a 5-mL balloon.15,18 This amount ensures even inflation of the balloon and prevents deflection of the catheter tip1 (Figure 2
). Larger balloons, however, such as the 30-mL size originally designed to decrease postoperative bleeding, result in greater impairment to bladder drainage15 because the eyes of the catheter tip sit higher in the bladder8 (Figure 2
).

View larger version (33K):
[in this window]
[in a new window]
|
Figure 2 Differences in residual volume of urine with urethral catheters with different balloon sizes (10 mL and 30 mL) and deflection of catheter tip with a catheter with an asymmetrical balloon.
|
|
Unfortunately, indwelling urinary catheters and urological procedures are a major source of urinary tract infections. For ICU patients, indwelling catheters should be removed when a patients condition is stable because a direct relationship exists between dwell time and incidence of infection.13 Additional risk factors for urinary tract infection are being female (because of the short urethra and its proximity to the anus) and extended ICU stay.19 In a recent Canadian multisite study,19 the incidence of ICU-acquired urinary tract infection was 9%. In a study20 of 112 medical ICUs in 97 hospitals across the United States, urinary tract infections were responsible for 31% of the nosocomial infections, and 95% of these were associated with urinary catheters. Connecting the catheter to an aseptic closed drainage system assists in reducing the incidence of infection,13 and the maintenance of a sterile continuously closed drainage system is recommended by the US Centers for Disease Control and Prevention.21
Sources of urinary tract infection can be both endogenous and exogenous.22 Although a major entry point is the periurethral area, additional entry points for infectious organisms are the juncture of the catheter and tubing and the collection container.13 Although not extensively studied, incomplete emptying of the bladder, defined as residual urine, also has been implicated as a factor in the development of urinary tract infections in catheterized patients.23 Despite continuous drainage when a catheter is in place, the remaining urine continually reinoculates the bladder with urethral and other organisms.24 This situation led to a recommendation that attempts be made to reduce the amount of residual urine to zero.23
 |
Implications of the Current Design of Indwelling Bladder Catheters for Patients Care
|
|---|
Probably the most common indwelling urinary catheter is the Foley catheter, originally designed by Frederick Foley in the 1930s.1 The standard Foley catheter generally has 2 lumens, 1 for drainage of urine and 1 for inflation of the balloon. Currently, each catheter has a solitary drainage hole near the tip, approximately 1.5 cm above the base of a 5-mL balloon. As a result, residual urine occurs as urine accumulates below the balloon13 (Figure 2
), leading to inaccurate measurements of urine output and possibly an increased predisposition to urinary tract infection.24,25 Increased residual urine occurs if larger balloon sizes are used15,25 or if balloons are overinflated to more than the recommended volume (Figure 2
). Although investigation has been limited, the results of preliminary studies with fluoroscopic imaging also suggest that for some patients, tenting of the catheter up the dome of the bladder may be the cause of incomplete urine evacuation26 (Figure 1
). Last, anatomical and physiological properties of the bladder, such as bladder shape or bladder sacculation and diverticula,27 also may play a role in residual urine.
Residual urine both with catheters in situ and immediately after catheter removal have been reported in several studies. In a study28 of 24 women of whom 11 had a urethral Foley catheter in place, a mean volume of 77 mL of urine remained in the bladder immediately after catheter removal. The authors28 reported that Foley catheters are relatively ineffective in completely emptying the bladder. In another study,26 this potential for incomplete bladder emptying, confirmed by fluoroscopy, also was noted. Although 70% of men (n = 116) and 73% of women (n = 148) had no residual urine when a catheter was in situ, mean true residual was 76 mL (SD 76 mL) for the remaining 30% of men and 52 mL (SD 36 mL) for the remaining 27% of women. A mean residual urine volume of 22 mL (SD 25 mL) not removed by catheterization also occurred in a later study29 of 95 ambulatory women. Last, incomplete emptying of the bladder by means of an indwelling catheter occurred in 25 (69%) of 36 catheterizations of 12 patients with spinal cord injury who were in the supine position.23
In a more recent study30 of the effect of urine flow rate on bladder temperature, cardiac surgery patients with indwelling bladder catheters had urine output measured every 2 minutes for 130 minutes in the postoperative period. One finding was that an increase in urine output occurred with changes in patients body positions. Most likely this increase was due to the increased drainage of residual urine from within the bladder. As can be seen in Figure 3
, patients had noticeable increases in urine output when repositioned from supine to side-lying or high Fowlers positions.30 Fluid remaining below the catheter balloon also is evident in the photograph of an indwelling catheter placed within the bladder of a female cadaver (Figure 4
).

View larger version (20K):
[in this window]
[in a new window]
|
Figure 3 Measurements of urine output at 2-minute intervals in a patient after cardiac surgery. A, Repositioning of the patient into a side-lying position at intervals 8 and 20 (arrows) resulted in increases in urine output. B, A change in the patients position to high Fowlers position at interval 28 (arrow) resulted in an increase in urine output.
|
|
The effects of residual urine in the bladder on accurate measurement of flow rate and on development of urinary tract infection are unknown. In critically ill patients in whom accurate measurements of urine output are essential, the urine remaining within the bladder leads to an inaccurate and misleading measurement of urine production. This inaccurate measurement is of particular concern for patients for whom renal function is problematic. As a result, unwarranted interventions such as diuretic therapy may be implemented. The additional concern of whether this residual urine contributes to the development of urinary tract infection requires further investigation.
These concerns indicate that a modification in the design of indwelling Foley catheters currently in use in North America is worthy of investigation. In fact, as early as the 1960s, the need for development of a retention catheter with an additional drainage eyelet in the shaft immediately below the balloon was proposed31 (Figure 5
). Compared with a standard catheter (Figure 6
), a catheter with an additional drainage outlet (Figure 7
) may eliminate or lessen residual urine volumes and subsequently the associated error in measurements of urine flow rate. Because residual urine within the bladder may predispose patients to urinary infection, reduction of this pool of urine may reduce the occurrence of this infection, particularly in patients who have catheters in place for considerable lengths of time. Interestingly, in a 1983 publication32 describing various catheters available on the market in the United Kingdom, a 30-mL balloon Pemberton catheter (Warne Surgical Products Ltd., Armagh, Northern Ireland) and a specialized Coudé (elbowed)-tip Roberts catheter (Bard International Ltd., Crawley, England), both with an additional drainage eyelet below the balloon to drain residual urine, were available. Although the 30-mL balloon Pemberton catheter is no longer manufactured, Willy Rüsch GmbH (Kernen, Germany) has a similar catheter, but again with a 30-mL balloon and only available in larger 18F to 24F size (C. Perry, written communication, August 23, 2002). In North America, a recently introduced 18F 4-way irrigation Coudé-tipped catheter with drainage eyelets below the balloon (Bard Medical Division, Covington, Ga) also has become available (T. J. Kelly, written communication, August 20, 2002). However, although specialized catheters generally used for urology patients exist, apparently no standard indwelling Foley catheter with a straight tip, a 5-mL balloon, and one or more drainage eyelets below the balloon that would be suitable for ICU patients is currently on the market.

View larger version (119K):
[in this window]
[in a new window]
|
Figure 6 Use of a standard urinary catheter with a solitary drainage opening above the balloon results in residual fluid within the container.
|
|

View larger version (113K):
[in this window]
[in a new window]
|
Figure 7 Use of a modified urinary catheter with an additional drainage hole below the balloon results in no residual fluid in the container.
|
|
Although various factors may play a role in the amount of urine remaining within the bladder in patients with a urinary catheter in place, clearly catheters play an important part. The oddity is that so little research has been done on this fundamental function of urinary catheters; the clear need for further exploration of this area has been noted by other authors.26 The question of why research on this topic is so sparse remains a concern.
Clinicians need to be involved in the selection of appropriate devices and/or be advocates for the development and testing of instruments and tools that provide data used in making treatment decisions in the clinical area. Thus, I recommend that the development and testing of a 5-mL balloon retention catheter with one or more additional drainage eyelets below the balloon be considered, particularly for patients in critical care, where monitoring of urine output is a frequent and necessary procedure. Because urine output is used as an indicator of renal function33 and accurate measurements of urine are essential in evaluating both fluid status and renal perfusion,6 the importance of a catheter that allows the least amount of urine to remain within the bladder is paramount. Subsequent comparisons of the efficacy of these newly designed catheters in draining urine with the efficacy of the indwelling catheters currently on the market should then ensue. In addition, the incidence of urinary tract infections in patients who have the currently available catheters in place should be compared with the incidence in patients who have the newly designed catheter.
 |
Acknowledgments
|
|---|
David Pinchuk, MD, and Charlene Shatsky, RN, BN, Victoria General Hospital, Winnipeg Manitoba, are acknowledged for their early review of this manuscript.
 |
References
|
|---|
- Pomfret IJ. Catheters: design, selection and management. Br J Nurs. 1996;5:245251.[Medline]
- Jain P, Parada JP, David A, Smith LG. Overuse of the indwelling urinary tract catheter in hospitalized medical patients. Arch Intern Med. 1995;155:14251429.[Abstract/Free Full Text]
- Kirklin JW, Barratt-Boyes BG. Cardiac Surgery. New York, NY: Churchill Livingstone; 1993.
- Thelan LA, Davie JK, Urden LD. Textbook of Critical Care Nursing: Diagnosis and Management. St Louis, Mo: Mosby; 1990.
- Frankenfield D, Cooney RN, Smith JS, Rowe WA. Age-related differences in the metabolic response to injury. J Trauma. 2000;48:4957.[Medline]
- Ledoux D, Luikart H. Cardiac surgery. In: Woods SL, Froelicher ES, Motzer SA, eds. Cardiac Nursing. 4th ed. Philadelphia, Pa: Lippincott Williams & Wilkins; 2000:580613.
- Spence AP. Basic Human Anatomy. 3rd ed. Redwoood City, Calif: Benjamin/Cummings; 1990.
- Parker LJ. Urinary catheter management: minimizing the risk of infection. Br J Nurs. 1999:8;563566, 568, 570, passim.[Medline]
- Damaser MS, Lehman SL. The effect of urinary bladder shape on its mechanics during filling. J Biomech. 1995;28:725732.[Medline]
- Stanton BA, Koeppen, BM. The kidney. In: Berne RM, Levy MN, eds. Physiology. 4th ed. St Louis, Mo: Mosby; 1998:675776.
- Constantinou CE, Djurhuus JC. Pyeloureteral dynamics in the intact and obstructed multicalyceal kidney. Am J Physiol. 1981;241:R398R411.[Medline]
- Haylen BT. Residual urine volumes in a normal female population: application of transvaginal ultrasound. Br J Urol. 1999;64:347349.
- Kunin CM. Urinary Tract Infections: Detection, Prevention, and Management. Baltimore, Md: Williams & Wilkins; 1997.
- Stoller ML. Retrograde instrumentation of the urinary tract. In: Tanagho EA, McAninch JW, eds. Smiths General Urology. 15th ed. New York, NY: McGraw-Hill; 2000:196207.
- Newman DK. Managing indwelling urethral catheters. Ostomy Wound Manage. December 1998;44:2628, 30, 32, passim.
- Coxon V. Inserting an indwelling catheter: female. In: Altman GB, Buschel P, Coxon V, eds. Delmars Fundamentals and Advanced Nursing Skills. Albany, NY: Delmar Thomson Learning; 2000:701708.
- McConnell EA. Inflating an indwelling urinary catheter balloon. Nursing 95. December 1995;25:13.[Medline]
- Robinson J. Urethral catheter selection. Nurs Stand. May 713, 2002;15:3942.
- Laupland KB, Zygun DA, Davies HD, Church DL, Louie TJ, Doig CJ. Incidence and risk factors for acquiring nosocomial urinary tract infection in the critically ill. J Crit Care. 2002;17:5057.[Medline]
- Richards MJ, Edwards JR, Culver DH, Gaynes RP. Nosocomial infections in medical intensive care units in the United States. Crit Care Med. 1999;27:887892.[Medline]
- Wong ES. Guidelines for prevention of catheter-associated urinary tract infections. Centers for Disease Control and Prevention. Available at: http://www.cdc.gov/ncidod/hip/GUIDE/uritract.htm. Accessed January 24, 2005.
- Tambyah PA, Halvorson KT, Maki DG. A prospective study of pathogenesis of catheter-associated urinary tract infections. Mayo Clin Proc. 1999;74:131136.[Abstract]
- Jensen AE, Hjeltnes N, Berstad J, Stanghelle JK. Residual urine following intermittent catheterisation in patients with spinal cord injuries. Paraplegia. 1995;33:693696.[Medline]
- Falkiner FR. The insertion and management of indwelling urethral catheters: minimizing the risk of infection. J Hosp Infect. 1993; 25:7990.[Medline]
- Getliffe K. Informed choices for long-term benefits: the management of catheters in continence care. Prof Nurse. 1993;9:122126.[Medline]
- Stoller ML, Millard RJ. The accuracy of catheterized residual urine. J Urol. 1989; 141:1516.[Medline]
- Purkiss SF. Assessment of residual urine in men following catheterisation. Br J Urol. 1990;66:279280.[Medline]
- Haylen BT, Frazer MI, MacDonald JH. Assessing the effectiveness of different urinary catheters in emptying the bladder: an application of transvaginal ultrasound. Br J Urol. 1989;64:353356.[Medline]
- Goode PS, Lochler JL, Bryant RL, Roth DL, Burgio KL. Measurement of postvoid residual urine with portable transabdominal bladder ultrasound scanner and urethral catheterization. Int Urogynecol J Pelvic Floor Dysfunct. 2000;11:296300.[Medline]
- Fallis WM. Core and Bladder Temperature Gradient in Critically Ill Adults: Urine Flow Rate as a Factor [doctoral dissertation]. Seattle, Wash: University of Washington; 2002.
- Pemberton M. Modified balloon catheter. Br Med J. 1967;1:485.
- Blannin J. Catheter guide. Community Outlook. September 15,1983:251254, 259263.
- Bellomo R, Chapman M, Finfer S, Hickling K, Myburgh J. Low-dose dopamine in patients with early renal dysfunction: a placebo-controlled randomised trial. Australian and New Zealand Intensive Care Society (ANZICS) Clinical Trials Group. Lancet. 2000;356:21392143.[Medline]