Critical Care Nurse. 2009;29: 67-73 doi:10.4037/ccn2009607
Copyright © 2009 by the American Association of Critical-Care Nurses.
This Article
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Baranchuk, A.
Right arrow Articles by Abdollah, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Baranchuk, A.
Right arrow Articles by Abdollah, H.


ECG

Electrocardiography Pitfalls and Artifacts: The 10 Commandments

Adrian Baranchuk, MD
Catherine Shaw, RCT
Haitham Alanazi, MD
Debra Campbell, RN, BScN, CCN(C)
Kathy Bally, RN, BNSc, CCN(C)
Damian P. Redfearn, MD, MB, ChB, MRCPI
Christopher S. Simpson, MD, FRCPC
Hoshiar Abdollah, MD, MB, ChB


All authors are in the Department of Cardiology at Queen’s University in Kingston, Ontario, Canada.

To purchase electronic or print reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. Phone, (800) 899-1712 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints{at}aacn.org.

dotmore
To learn more about ECG algorithms, read "ECG Computer Algorithms," by Michele M. Pelter and Mary G. Carey in the American Journal of Critical Care 2008;17:581–582. Available at www.ajcconline.org.

eLetters
Now that you’ve read the article, create or contribute to an online discussion about this topic using eLetters. Just visit www.ccnonline.org and click "Respond to This Article" in either the full-text or PDF view of the article.

Financial Disclosures
None reported.

Corresponding author: Adrian Baranchuk, MD, Cardiac Electrophysiology and Pacing, Kingston General Hospital, Queen’s University, Kingston, Ontario, Canada K7L 2V7 (e-mail: barancha{at}KGH.KARI.NET).


Many potential pitfalls can adversely affect the interpretation of 12-lead ambulatory and telemetry electrocardiograms (ECGs). Artifacts, for example, are a common finding in patients who require ECG monitoring. Artifacts are defined as ECG abnormalities that may be due to sources other than the electrical activity of the heart. Failure to correctly distinguish between an arrhythmia and artifact can result in misdiagnosis and unnecessary therapeutic interventions.1

The most common causes of artifacts originate from internal (physiological) and external (non-physiological) sources (Table 1Go). Artifacts created from these sources can simulate arrhythmias such as atrial flutter and ventricular tachycardia.2


View this table:
[in this window]
[in a new window]

 
Table 1 Most frequent causes of electrocardiographic artifact and other pitfalls

 
Electrode misplacement is another common artifact. Such misplacement may lead to changes in ECG morphology that could potentially be interpreted as ischemic in origin.3 Electrode misplacements can also mimic serious arrhythmias and lead to misdirected therapeutic decisions.4 Electrode misplacement is a relatively frequent finding in ECGs done in outpatient clinics (0.4%) and is even more common in intensive care units (4%).4 The correct position for precordial ECG electrodes is illustrated in Figure 1Go.


Figure 1
View larger version (105K):
[in this window]
[in a new window]

 
Figure 1 A, Standard position of precordial electrodes. B, Normal 12-lead electrocardiogram.

 
Several telltale clues can help clinicians identify potential signs of electrode misplacements and artifacts. In this article, we introduce an algorithm that we developed to assist nurses and physicians in rapidly recognizing those clues and review 10 of the most common ECG pitfalls and artifacts.

Algorithm to Identify ECG Electrode Misplacements or Artifacts (REVERSE)

The indicators of electrode misplacements or artifacts that clinicians need to look for can be easily remembered by using the mnemonic REVERSE (Table 2Go). With this mnemonic in mind, careful and systematic examination of ECGs will help rule out problems with the recording.


View this table:
[in this window]
[in a new window]

 
Table 2 REVERSE mnemonic: an easy approach to remember the most frequent causes of electrocardiographic artifacts and pitfalls

 
We describe the 10 most common ECG pitfalls and artifacts seen in our practice, presented in the form of commandments. All the ECG examples provided for this review were run at 25 mm/s, 10 mm/mV, and 100 Hz.

ECG Pitfalls and Artifacts: The 10 Commandments

1. You shall not reverse the electrodes: reversal of left arm and right arm electrodes
Reversing the electrodes is one of the most common errors made when placing the ECG on a patient. Such reversal produces leads I and AVL with reverse polarity of all normal deflections (negative P wave, QRS complex, and T wave). In addition, polarity is reversed in lead AVR (positive P and QRS; Figure 2Go). The differential diagnosis is dextrocardia (the heart is positioned on the right side). In dextrocardia, however, the progression of the R wave in pre-cordial leads is reversed, whereas with electrode reversal, the progression is normal.5


Figure 2
View larger version (78K):
[in this window]
[in a new window]

 
Figure 2 Reversal of electrodes for left arm and right arm. Note reverse polarity (arrows) in leads I and AVL (negative P, QRS, and T deflections). Note also reverse polarity in AVR (positive P and QRS deflections).

 
2. You shall not treat the ECG, treat the patient: artifact mimicking ventricular tachycardia
The possibility of tremor or other interference inducing an artifact that mimics ventricular tachycardia should be considered when the ECG does not match the patient’s clinical findings. A normal heart rate obtained by pulse or auscultation in an asymptomatic patient at the same time the ECG shows apparent ventricular tachycardia confirms the diagnosis. Reduction of the tremor by holding the limb or placing the electrodes on the torso will reduce interference.2,6,7 "Tracking" the R-R intervals is helpful if they can be identified before the pseudo– ventricular tachycardia. Look for R-R intervals that continue into the wide complex rhythm to see the presence of normal ventricular depolarizations throughout the pseudo–ventricular tachycardia. With careful measuring to see where normal beats should be, they will often "jump out" at the observer and become obvious, whereas at first glance they may be completely obscured. Pseudo–ventricular tachycardia has 3 characteristic signs8 (Figure 3Go):


Figure 3
View larger version (73K):
[in this window]
[in a new window]

 
Figure 3 Artifact that mimics ventricular tachycardia. Note the "sinus" sign (asterisk) in lead I (a P wave is visible before the second QRS complex). The "spike" and "notch" signs (arrows) can be seen by tracking the previous R-R interval among the wide-QRS-like complexes.

 
  1. Sinus sign: one of the frontal leads (I, II, or III) shows normal P waves, QRS complexes, and T waves because usually one of the upper limbs is free of tremor or movement.
  2. Spike sign: tiny spikes can be seen among wide QRS-like complexes.
  3. Notch sign: notches are superimposed in the wide QRS-like complex artifact that "time out" with preceding R-R intervals.

3. You shall not reverse the electrodes: reversal of left arm and left leg electrodes
Amplitude of the P wave in lead I greater than in lead II and/or P-wave terminal positive component in lead III (Abdollah sign) will confirm reversal of the left arm and left leg leads.9 Confirmation with a second ECG is usually required (Figure 4Go).


Figure 4
View larger version (64K):
[in this window]
[in a new window]

 
Figure 4 Reversal of left arm and left leg electrodes. Note that the amplitude of the P wave in lead I is greater than that in lead II (arrows), and the P wave in lead III has a small positive terminal component (Abdollah sign).

 
4. You shall not reverse the electrodes: reversal of precordial leads (V1 and V6 )
The most common reversal of the precordial leads is an exchange of V1 and V6. The way to recognize this problem is by assessing the R-wave progression in the precordial leads. Normally, the R wave will increase its amplitude from V1 to V6 and the S wave will decrease its amplitude. In the reversal situation, a tall R wave can be seen in V1 and a deep S wave in V61013 Potential diagnostic misinterpretations include right bundle branch block, old posterior myocardial infarction, right ventricular hypertrophy, and left-sided accessory pathways (Figure 5Go).


Figure 5
View larger version (50K):
[in this window]
[in a new window]

 
Figure 5 Reversal of precordial electrodes. Tall R wave is seen in V1, and small R wave and deep S wave (arrows) are seen in V6, indicating V1 to V6 electrode reversal.

 
5. You shall check if the patient is calm and quiet: tremor
Tremor-induced artifact may mimic supraventricular arrhythmias (atrial flutter/atrial fibrillation) or if the artifact has sufficient amplitude, it can also mimic ventricular tachycardia and ventricular fibrillation. The correct diagnosis can be made on the basis of simple observations such as the presence of the pseudoarrhythmia when the patient moves (tremor). Careful analysis may reveal discrete components of the QRS complexes (matching the previous R-R intervals if present) through the pseudoarrhythmia ("notches sign").8,1416 Misinterpretation of tremor-induced artifact may lead to serious medical errors such as the initiation of long-term use of anticoagulants for pseudo– atrial fibrillation.17 Figure 6Go shows pseudo–atrial flutter that was a tremor-induced artifact.


Figure 6
View larger version (75K):
[in this window]
[in a new window]

 
Figure 6 Tremor-induced artifact. Note pseudo–flutter waves (arrowheads) in the inferior leads. Normal P waves (arrow) are visible in the rest of the leads.

 
6. You shall turn off your cell phone: electromagnetic interference
Electromagnetic interference (EMI) with medical devices by cell phones is a well-recognized problem.1821 Even though considerable controversy remains about the use of cell phones in hospitals, the evidence is clear that cell phones can produce EMI with many different medical devices (eg, ECG monitor, ventilator, infusion pump, dialysis machine, apnea monitor, external pacemaker, internal pacemaker, and defibrillator). The ability of a cell phone or a wireless device to induce EMI depends on the distance, the ability of medical equipment to resist EMI, and the technology of the cell phone (digital vs analog, which are the 2 basic systems cell phones use to operate, and single-band of operation vs dual and frequency band of operation).19 As shown by previous investigators,19 a 1-m (3.28 ft) distance between the source of EMI and medical devices safely eliminates EMI. Only a few cases in which cell phones and wireless devices interfered with ECG machines have been reported.20 We simulated a case created in our laboratory by activating a cell phone (digital) less than 25 cm (9.8 in) from the ECG machine acquisition module (MAC 5000 Resting ECG Analysis System, GE Medical Systems, Waukesha, Wisconsin). The rapid, sharp, and low-amplitude signals disappeared when the cell phone was removed or deactivated (Figure 7Go). ECG technicians and nurses should avoid using cell phones when they are recording ECGs until further research in this area is available. This limitation may also have implications for paramedics and ambulance attendants who obtain and interpret ECGs on patients in the field.


Figure 7
View larger version (76K):
[in this window]
[in a new window]

 
Figure 7 Electromagnetic interference. Sharp and fast spikes (arrowheads) were induced by a digital cell phone activated less than 25 cm from the acquisition module of the electrocardiography machine.

 
7. You shall know where the limbs are: electrodes placed on the torso
Electrodes are placed on the torso near the extremities rather than on the limbs for different reasons. During an emergency, placing leads on the torso reduces the time needed for undressing the patient and in most cases will allow a correct ECG diagnosis.22 However, in most circumstances, the torso position should not replace the standard position on the limbs. The torso position induces a change in how the electrical vectors are recorded. Pseudo–Q waves and pseudo–ST-segment elevation in the inferior leads5,23,24 could potentially be misinterpreted as myocardial infarction (Figure 8Go).


Figure 8
View larger version (126K):
[in this window]
[in a new window]

 
Figure 8 Torso position of the electrodes. A, Baseline. B, Torso position. Note fast Q waves and ST-segment elevation in the inferior leads.

 
8. You shall not place telemetry electrodes on top of ECG electrodes: telemetry interference
Placing the telemetry electrodes on top of the ECG electrodes or vice versa is a common mistake. Usually, telemetry electrodes are placed in the same region where the ECG electrodes need to be placed. This superimposition of electrodes may create a distortion of the ST segment that mimics ST-segment elevation or arrhythmias due to EMI of the telemetry on the ECG machine25 (Figure 9Go).


Figure 9
View larger version (78K):
[in this window]
[in a new window]

 
Figure 9 Telemetry interference. Telemetry electrodes were placed on top of the electrocardiographic electrodes. Note pseudo–ST-segment elevation (arrows) in leads V2 and V6.

 
9. You shall not reverse the electrodes: reversal of right leg and left arm or right arm electrodes
If a reversal involves the right leg and one of the arms, the recording will be zero potential difference between the legs.5 This pseudoasystole in an isolated lead may occur in lead II (reversal of right arm and right leg electrodes) or in lead III (reversal of left arm and right leg electrodes; Figure 10Go).


Figure 10
View larger version (66K):
[in this window]
[in a new window]

 
Figure 10 Reversal of right arm and right leg electrodes. Note pseudoasystole (arrow) in lead III.

 
10. The patient should be lying down, calm and relaxed (if possible): ECG done with patient sitting at 90°
In some clinical situations (eg, decompensated heart failure, respiratory insufficiency, orthopedic limitations), the ECG must be recorded with the patient sitting upright or in a semi-Fowler’s position. Changing the body position can affect the QRS axis and QRS amplitude.26,27 Currently, no distinctions in the recording methods need to be made when an ECG is recorded with the patient sitting (~90°); however, recognizing slight alterations of the QRS complex may be helpful to avoid wrong interpretations. An annotation indicating the position of the patient (if different than usual) may be helpful for the physician interpreting the recordings. In this case, note the reduction of the QRS amplitude in lead III, which is a lead that is particularly sensitive to changes in diaphragmatic position (Figure 11Go).


Figure 11
View larger version (64K):
[in this window]
[in a new window]

 
Figure 11 Electrocardiogram recorded with patient sitting upright at 90°. Note low-amplitude QRS complex (arrows) in lead III.

 
Conclusions

The ECG is one of the most valuable tools in our daily practice. Many health care providers interpret ECGs and initiate therapeutic interventions on the basis of such interpretations. Recognizing ECG artifacts and other pitfalls will enable clinicians to avoid unnecessary therapeutic interventions and may allow them to correct the recording methods to obtain a proper ECG.

PRIME POINTS

References

  1. Chase C, Brady WJ. Artifactual electrocardiographic change mimicking clinical abnormality on the ECG. Am J Emerg Med. 2000;18(3):312–316.[CrossRef][Medline]
  2. Srikureja W, Darbar D, Reeder GS. Tremor-induced ECG artifact mimicking ventricular tachycardia. Circulation. 2000;102(11): 1337–1338.[Free Full Text]
  3. Rudiger A, Schöb L, Follath F. Influence of electrode misplacement on the electrocardiographic signs of inferior myocardial ischemia. Am J Emerg Med. 2003;21(7):574–577.[CrossRef][Medline]
  4. Rudiger A, Hellermann JP, Mukherjee R, et al. Electrocardiographic artifacts due to electrode misplacement and their frequency in different clinical settings. Am J Emerg Med. 2007;25(2):174–178.[CrossRef][Medline]
  5. Drew BJ. Pitfalls and artifacts in electrocardiography. Cardiol Clin. 2006;24(3):309–315.[CrossRef][Medline]
  6. Knight BP, Pelosi F, Michaud GF, Strickberger SA, Morady F. Physician interpretation of electrocardiographic artifact that mimics ventricular tachycardia. Am J Med. 2001;110(5):335–338.[CrossRef][Medline]
  7. Knight BP, Pelosi F, Michaud GF, Strickberger SA, Morady F. Clinical consequences of electrocardiographic artifact mimicking ventricular tachycardia. N Engl J Med. 1999; 341(17):1270–1274.[Free Full Text]
  8. Huang CY, Shan DE, Lai CH, et al. An accurate electrocardiographic algorithm for differentiation of tremor-induced pseudo-ventricular tachycardia and true ventricular tachycardia. Int J Cardiol. 2006;111(1): 163–165.[CrossRef][Medline]
  9. Abdollah H, Milliken JA. Recognition of electrocardiographic left arm/left leg lead reversal. Am J Cardiol. 1997;80(9):1247–1249.[CrossRef][Medline]
  10. Wenger W, Kligfield P. Variability of precordial electrode placement during routine electrocardiography. J Electrocardiol. 1996; 29(3):179–184.[Medline]
  11. Peberdy MA, Ornato JP. Recognition of electrocardiographic lead misplacements. Am J Emerg Med. 1993;11(4):403–405.[CrossRef][Medline]
  12. Marafioti V, Variola A. Pseudoinfarction pattern by misplacement of electrocardiographic precordial leads. Am J Emerg Med. 2004;22(1):62–63.[CrossRef][Medline]
  13. Hurst JW. "Switched" precordial leads. Circulation. 2000;101(24):2870–2871.[Free Full Text]
  14. Llinas R, Henderson VH. Tremor as a cause of pseudo-ventricular tachycardia. N Engl J Med. 1999;341:1275.[Free Full Text]
  15. Bhatia L, Turner DR. Parkinson’s tremor mimicking ventricular tachycardia. Age Ageing. 2005;34(4):410–411.[Abstract/Free Full Text]
  16. Ortega-Carnicer J. Tremor-related artifact mimicking ventricular tachycardia. Resuscitation. 2005;65(3):243–244.[CrossRef][Medline]
  17. Finsterer J, Stollberger C, Gatterer E. Oral anticoagulation for ECG tremor artifact simulating atrial fibrillation. Acta Cardiol. 2003; 58(5):425–429.[CrossRef][Medline]
  18. Morrissey JJ, Swicord M, Balzano Q. Characterization of electromagnetic interference of medical devices in the hospital due to cell phones. Health Phys. 2002;82(1):45–51.[CrossRef][Medline]
  19. Lawrentschuk N, Bolton DM. Mobile phone interference with medical equipment and its clinical relevance: a systematic review. Med J Aust. 2004;181(3):145–149.[Medline]
  20. Baranchuk A, Kang J, Shaw C, et al. Electromagnetic interference of communication devices on ECG machines. Clin Cardiol. In press.
  21. Klein AA, Djaiani GN. Mobile phones in the hospital—past, present and future. Anaesthesia. 2003;58(4):353–357.[CrossRef][Medline]
  22. Takuma K, Hori S, Sasaki J, et al. An alternative limb lead system for electrocardiographs in emergency patients. Am J Emerg Med. 1995;13(5):514–517.[CrossRef][Medline]
  23. Jowett NI, Turner AM, Cole A, Jones PA. Modified electrode placement must be recorded when performing 12-lead electrocardiograms. Postgrad Med J. 2005;81(952):122–125.[Abstract/Free Full Text]
  24. Kligfield P, Gettes LS, Baileiy JJ, et al. Recommendations for the standardization and interpretation of the electrocardiogram, I: the electrocardiogram and its technology. Circulation. 2007;115(10):1306–1324.[Abstract/Free Full Text]
  25. Mirvis DM, Berson AS, Goldberger AL, et al. Instrumentation and practice standards for electrocardiographic monitoring in special care units. A Report for Health Professionals by a Task Force of the Council on Clinical Cardiology, American Heart Association. Circulation. 1989;79(2):464–471.[Abstract/Free Full Text]
  26. Nelwan SP, Meij SH, van Dam TB, Kors JA. Correction of ECG variations caused by body position changes and electrode placement during ST-T monitoring. J Electrocardiol. 2001;34(suppl):213–216.[CrossRef][Medline]
  27. Madias JE. Comparability of the standing and supine standard electrocardiograms and standing sitting and supine stress electrocardiograms. J Electrocardiol. 2006;39(2):142–149.[CrossRef][Medline]




This Article
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar